Back to all papers

A Unified Model for Highly Accurate ECG-Free Dynamic Coronary Roadmapping Using Spatio-Temporal Transformers

July 9, 2026arxiv logopreprint

Authors

Saahil Islam,Sebastian Piat,Venkatesh N. Murthy,Serkan Cimen,Puneet Sharma,Andreas Maier,Florin C. Ghesu

Abstract

Percutaneous Coronary Intervention (PCI) is a minimally invasive procedure used to restore coronary blood flow obstructed by atherosclerotic plaque. During PCI, repeated injections of iodine-based contrast agents are required to visualize the coronary arteries and guide interventional devices. However, frequent contrast injections increase radiation exposure and the risk of contrast-induced nephropathy, with acute kidney injury reported in up to 30% of patients with renal impairment. Dynamic Coronary Roadmapping (DRM) reduces these risks by overlaying a precomputed angiographic vessel map onto live fluoroscopy and continuously updating it throughout the procedure. Accurate DRM relies on precise cardiac phase matching between angiography and fluoroscopy, together with reliable catheter tip tracking for motion compensation. These tasks remain challenging in ECG-free settings and when only limited manual annotations are available. We present a unified DRM framework that simultaneously performs cardiac phase matching and catheter tip tracking for accurate real-time guidance. Our method employs a large-scale spatio-temporal encoder pretrained on 16 million X-ray frames to learn cardiac motion dynamics. To the best of our knowledge, this is the first application of large-scale spatio-temporal pretraining for motion compensation in DRM. We further introduce auxiliary tasks based on ECG R-peak detection and catheter tip tracking, improving optimization while eliminating the need for extensive catheter mask annotations. Finally, a majority-voting postprocessing strategy aggregates temporal predictions, improving robustness and providing a confidence score that correlates with phase-matching error. Comprehensive evaluation on clinical X-ray datasets demonstrates state-of-the-art performance, achieving low temporal misalignment and robust phase-matching accuracy suitable for real-time DRM.

Topics

eess.IV

Ready to Sharpen Your Edge?

Subscribe to join 11k+ peers who rely on RadAI Slice. Get the essential weekly briefing that empowers you to navigate the future of radiology.

We respect your privacy. Unsubscribe at any time.