Intended Use

Image processing software providing quantitative and qualitative analysis from previously acquired CT DICOM images for evaluation and assessment of cardiovascular diseases. Functions include heart segmentation and volume measurement, coronary calcium quantification, aorta segmentation, aortic diameter measurement, and threshold-based highlighting of enlarged diameters.

Technology

Software-only image post-processing application that uses deep learning algorithms to analyze CT chest data. It segments the heart and aorta, detects and quantifies coronary calcium, identifies anatomical landmarks on the aorta for diameter measurements according to AHA guidelines, and classifies measurements with threshold-based categories. Validated on non-cardiac chest CT from multiple CT vendors.

Performance

Performance was evaluated by software validation including non-clinical bench testing and clinical data-based software validation. The software demonstrated conformity with FDA consensus standards (DICOM, IEC 62304, ISO 14971). Clinical validation with retrospective studies showed high correlation (0.96) for coronary calcium volume compared to predicate device and average absolute aorta diameter measurement error of 1.6 mm. Testing supported claims of substantial equivalence and safety.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    11/23/2018

    9 months
  • 2

    FDA Approval

    9/10/2019

Ready to Sharpen Your Edge?