Photon-counting detector CT of the brain reduces variability of Hounsfield units and has a mean offset compared with energy-integrating detector CT.
Authors
Affiliations (1)
Affiliations (1)
- From the Department of Diagnostic and Interventional Radiology, Universityhospital Freiburg, Germany (TS, FL, SR, MFR, TS, AF, JW, FB), Medical Physics, Universityhospital Freiburg, Germany (MR, EK), Department of Stereotactic and Functional Neurosurgery, Universityhospital Freiburg, Germany (MR), and the Department of Neuroradiology, Universityhospital Freiburg, Germany (HU, AR).
Abstract
Distinguishing gray matter (GM) from white matter (WM) is essential for CT of the brain. The recently established photon-counting detector CT (PCD-CT) technology employs a novel detection technique that might allow more precise measurement of tissue attenuation for an improved delineation of attenuation values (Hounsfield units - HU) and improved image quality in comparison with energy-integrating detector CT (EID-CT). To investigate this, we compared HU, GM vs. WM contrast, and image noise using automated deep learning-based brain segmentations. We retrospectively included patients who received either PCD-CT or EID-CT and did not display a cerebral pathology. A deep learning-based segmentation of the GM and WM was used to extract HU. From this, the gray-to-white ratio and contrast-to-noise ratio were calculated. We included 329 patients with EID-CT (mean age 59.8 ± 20.2 years) and 180 with PCD-CT (mean age 64.7 ± 16.5 years). GM and WM showed significantly lower HU in PCD-CT (GM: 40.4 ± 2.2 HU; WM: 33.4 ± 1.5 HU) compared to EID-CT (GM: 45.1 ± 1.6 HU; WM: 37.4 ± 1.6 HU, p < .001). Standard deviations of HU were also lower in PCD-CT (GM and WM both p < .001) and contrast-tonoise ratio was significantly higher in PCD-CT compared to EID-CT (p < .001). Gray-to-white matter ratios were not significantly different across both modalities (p > .99). In an age-matched subset (n = 157 patients from both cohorts), all findings were replicated. This comprehensive comparison of HU in cerebral gray and white matter revealed substantially reduced image noise and an average offset with lower HU in PCD-CT while the ratio between GM and WM remained constant. The potential need to adapt windowing presets based on this finding should be investigated in future studies. CNR = Contrast-to-Noise Ratio; CTDIvol = Volume Computed Tomography Dose Index; EID = Energy-Integrating Detector; GWR = Gray-to-White Matter Ratio; HU = Hounsfield Units; PCD = Photon-Counting Detector; ROI = Region of Interest; VMI = Virtual Monoenergetic Images.