Geometry Calibration in Tomography with a Differentiable Ray-Based Model

June 19, 2026arxiv logopreprint

Authors

Youssef Haouchat,Aleix Boquet-Pujadas,Sepand Kashani,Philippe Thévenaz,Michael Unser

Abstract

Geometric misalignments between the nominal and true acquisition parameters in tomography degrade reconstructions. We propose a framework that jointly reconstructs the volume and calibrates the acquisition geometry for arbitrary source--detector configurations. The core of our framework is an x-ray transform operator whose gradients with respect to the acquisition geometry can be efficiently computed with a ray-tracing method of structure and computational complexity similar to those of the forward operator. We represent the volume in a B-spline basis to provide a continuously differentiable model. This results in a better-behaved optimization landscape compared to voxel-based representations. We validate our framework with CT, micro-CT, nano-CT, and positron emission tomography data under a variety of geometric misalignments.