Intended Use

Intended for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope.

Technology

The devices are single-use, sterile (Ethylene Oxide) aspiration needles designed to work with ultrasound endoscopes. Features include adjustable sheath lengths, specific needle gauges (21G and 22G), working length of 700mm, ultrasound visualization compatibility, and luer lock connectors.

Performance

Performance testing included insertion, piercing, ultrasound visibility, needle extraction/retraction, aspiration performance, and mechanical strength tests, all meeting acceptance criteria. Biocompatibility testing confirmed the device is safe for intended use. Sterilization testing validated endotoxin limits and sterility assurance; shelf-life testing confirmed a 3-year shelf life.

Predicate Devices

No predicate devices specified