Rapid MLS is an AI-powered software that analyzes non-contrast CT scans to measure and visually display brain midline shifts. This helps clinicians assess conditions like intracranial hemorrhage and traumatic brain injury by providing quantitative and annotated images, supporting but not replacing clinical decision-making.
The Rapid MLS software device is designed to measure the midline shift of the brain from a NCCT acquisition and report the measurements. Rapid MLS analyzes adult cases using machine learning algorithms to identify locations and measurements of the expected brain midline and any shift which may have occurred. The Rapid MLS device provides the user with annotated images showing measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of NCCT cases.
Rapid MLS is a radiological computer-assisted image processing software using AI/ML algorithms within an integrated platform to analyze non-contrast CT images in DICOM format. It outputs annotated images showing differences between expected and measured brain midline, plus a JSON report with quantitative measurements.
The device was validated using 153 NCCT cases with ground truth established by 3 experts, showing a mean absolute error (MAE) of 0.7 mm versus experts' 1.0 mm, demonstrating non-inferiority. Validation included diverse scanner manufacturers and multiple clinical sites, confirming accurate performance under clinically relevant conditions.
No predicate devices specified
Submission
10/30/2024
FDA Approval
5/28/2025