Intended Use

Software device for quantitative and qualitative analysis of angiography DICOM data to support functional evaluation of coronary artery disease.

Technology

The system uses standard DICOM angiographic images from X-ray imaging in the cath lab, constructs a 3D computer model of coronary arteries, and simulates blood flow to compute FFRangio, a derived physiological metric. Operated on a mobile cart with computer processing and user interface, it allows real-time analysis during procedures.

Performance

Performance testing included software verification and validation, hardware safety testing, electromagnetic compatibility, transportation, and extensive clinical validation in multicenter international trials comparing FFRangio results with invasive FFR measurements. The pivotal FAST-FFR study demonstrated high sensitivity (93.5%) and specificity (91.2%) indicating accurate, reliable functional assessment of coronary artery disease. Human factors studies confirmed usability and safety for intended users.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    8/8/2018

    4 months
  • 2

    FDA Approval

    12/19/2018

Other devices from CathWorks Ltd