MeVis AVM Contour is a medical imaging software designed for qualified healthcare professionals to assist in manually creating target contours in the brain by registering 2D digital subtraction angiography images onto 3D CT scans. It supports treatment planning for arteriovenous malformations in stereotactic radiosurgery by exporting contours used in radiosurgery or radiotherapy planning systems.
Intended for medical professionals in neurosurgery or radiotherapy planning to manually create target contours for cranial regions by semi-automatic registration of 2D DSA images onto 3D CT scans for radiosurgery/-therapy treatment planning of arteriovenous malformations.
Software tool integrating manual contouring, semi-automatic 2D-3D image registration using X-ray, 2D DSA, and 3D CT images of the brain, and export of 3D CT-based structures as DICOM RTStruct for radiosurgery planning.
Validated by software verification (unit, functional, acceptance testing), usability testing, security testing, and bench tests demonstrating semi-automatic image registration accuracy of 0.44 mm aligning 3D CT to 2D DSA images, meeting or exceeding predicate device accuracy.
No predicate devices specified