Intended Use

The Swoop® Portable MR Imaging® System (V2) is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Technology

The device is a portable, ultra-low field MRI with a permanent magnet of approximately 65 mT. It includes an off-the-shelf user interface device for scan setup and execution, provides access to various MRI sequences (T1W, T2W, FLAIR, DWI/ADC), and uses a deep learning image reconstruction algorithm called Advanced Reconstruction to enhance image quality relative to linear reconstruction methods.

Performance

Performance testing included verification and validation according to international standards (e.g., IEC 62304 for software, IEC 60601 series for safety), biocompatibility of patient-contacting materials, and cleaning/disinfection validation. Advanced Reconstruction was evaluated with a large independent dataset for robustness and image quality, showing improved metrics (NMSE, SSIM, contrast-to-noise ratio) and positive radiologist image quality assessments relative to linear reconstruction, without loss of diagnostic consistency or unexpected artifacts.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    1/27/2025

    4 months
  • 2

    FDA Approval

    5/30/2025