Intended Use

Mini-X is intended for use by trained medical professionals for fluoroscopic imaging and guidance during diagnostic radiography and surgical procedures of the extremities, excluding neonates and mammography.

Technology

The system is a portable mobile imaging system that acquires, processes, and displays fluoroscopic images using a low-powered mono-block generator and a dynamic flat-panel detector. The device operates from a standard 120VAC outlet and incorporates sophisticated software for image processing and display, with capabilities up to 30 frames per second.

Performance

Bench testing including phantom image analysis confirmed that the Mini-X performs equivalently to predicate devices in image quality, spatial resolution, and radiation safety. The device passed electrical, mechanical, and radiation safety tests and complies with relevant IEC and FDA standards. Software validation followed FDA guidance and cybersecurity considerations were addressed.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    9/10/2024

    8 months
  • 2

    FDA Approval

    5/16/2025