Intended Use

Cornaris Intravascular Imaging System with Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. LumenCross Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.

Technology

The system consists of a cart-mounted computer and imaging engine with two models (P80-E and Mobile-E). It uses an internal rotating fiber optic imaging core to acquire OCT images via imaging catheter. The catheter includes a guidewire lumen and hydrophilic coating for intravascular navigation.

Performance

Performance testing included scan range, axial resolution, frame rate, sensitivity, dynamic range, pullback time, and fiber optic joint performance. The catheter was tested visually and dimensionally, and subjected to biological safety testing including cytotoxicity, sensitization, and in vivo thrombogenicity. Animal studies demonstrated safety and equivalence to predicate devices. No clinical study included.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    7/18/2024

    8 months
  • 2

    FDA Approval

    4/11/2025