Intended Use

Viewing, post-processing, qualitative and quantitative evaluation of blood vessels and cardiovascular MR images in DICOM format.

Technology

Myomics uses AI-powered semi-automatic segmentation algorithms to analyze cardiac MR images from 3.0 T scanners, with modules trained and validated on large datasets. It operates as standalone or integrated software and complies with DICOM and other medical imaging standards.

Performance

Bench testing confirmed functional equivalence with predicate devices including image import, display, contour drawing and semi-automatic segmentation. AI performance testing used 728 anonymized MR cases from multiple vendors with all modules achieving DICE scores above 0.7, meeting predefined acceptance criteria.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    7/1/2024

    8 months
  • 2

    FDA Approval

    2/28/2025