Intended Use

MammoScreen BD is a software application intended for use with compatible full field digital mammography and digital breast tomosynthesis systems, providing ACR BI-RADS 5th Edition breast density category assessment in asymptomatic women aged 40 and older undergoing screening mammography.

Technology

MammoScreen BD is a software-only medical device using AI deep learning algorithms to analyze mammograms and tomosynthesis images to output breast density classification per ACR BI-RADS 5th Edition, operating on DICOM image inputs and delivering results in JSON format or DICOM Structured Reports compatible with clinical PACS and reporting software.

Performance

Performance testing on retrospective mammography datasets (922 exams from U.S. and France sites) measured accuracy, agreement with radiologist consensus, and reproducibility across imaging views and modalities, achieving high weighted Cohen's kappa scores (~89 for four-category and ~84.5 for binary breast density classification). Subgroup analyses by image modality, patient age, breast thickness, and race showed consistent performance without bias. The software complies with IEC 62304 lifecycle standards and demonstrated non-inferiority to predicate device.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    5/31/2024

    4 months
  • 2

    FDA Approval

    10/2/2024

Other devices from Therapixel