Intended Use

Software for diagnosis and surgical planning of foot and ankle orthopedic conditions using X-ray imaging input.

Technology

Uses machine learning algorithms to generate 3D axes and mathematical models of foot and ankle anatomy and orthopedic fixation devices from weight-bearing X-ray images, producing automated case reports for surgical planning.

Performance

Performance testing included software verification, accuracy assessment against ground truth from expert clinicians, and validation on diverse X-ray and CBCT DRR datasets. Results met acceptance criteria of 95% model conformance within 1.0mm and angular accuracy ±1 degree, showing the device performs as intended without safety or effectiveness concerns.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    3/18/2024

    3 months
  • 2

    FDA Approval

    7/2/2024