Intended Use

Diagnostic ultrasound imaging and fluid flow analysis of the human body, including cardiac adult and various other anatomical indications.

Technology

The system uses diagnostic ultrasound imaging technology with pulsed doppler modes and an AI-based Smart Doppler View ID software feature. The AI algorithm analyzes acquired Doppler spectrums to automatically highlight the corresponding Doppler Calculation Package group on the touchscreen, assisting navigation and workflow but not performing measurements. The feature supports various transducers and runs on EPIQ and Affiniti systems with software version 11.0 or higher.

Performance

Non-clinical verification tests and software validation were performed, showing the AI algorithm achieved 97.5% accuracy (95% CI 96.3%-98.3%, p < 0.0001) in matching user-verified ground truth touchscreens for Doppler measurements using a clinical dataset of 1100 clips from 400 patients with diverse demographics. No new hardware or measurements were introduced, only workflow automation.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    11/28/2023

    2 months
  • 2

    FDA Approval

    2/13/2024