Intended Use

Qualitative and quantitative evaluation of cardiovascular MR images in DICOM format, providing measurements of 2D LV myocardial function for diagnostic information in patients with suspected heart disease.

Technology

Software analyzes CMR images via 2D feature tracking on cine SSFP images using a nearly incompressible deformable model. It calculates various deformation metrics including strain, strain rate, displacement, velocity, and torsion in radial, circumferential, and longitudinal directions without the use of AI or machine learning.

Performance

The software underwent verification and validation testing per ISO and FDA guidance standards. Testing included the use of analytical and realistic phantoms, comparisons with manual expert tracking, and adherence to ISO 13485, IEC 62304, ISO 14971, and DICOM standards. No clinical studies were required for substantial equivalence.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    8/31/2023

    3 months
  • 2

    FDA Approval

    12/7/2023