Intended Use

The Thirona LungQ software provides reproducible CT values for pulmonary tissue essential for quantitative support in diagnosis and follow up of pulmonary abnormalities from CT thoracic datasets.

Technology

LungQ is a stand-alone command-line software that performs 3D segmentation and isolation of lung sub-compartments, volumetric and density analyses, fissure evaluation, and generates reports from CT datasets. It operates on Linux and supports DICOM data input.

Performance

Performance testing included compliance with FDA-recognized consensus standards, software verification and validation, human factors validation, and a substantial equivalence study comparing LungQ v3.0.0 with predicate LungQ v1.1.0. Results demonstrated equivalence in lung volume and density measurements, fissure evaluation, and overall safety and effectiveness without clinical testing.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    8/10/2023

    5 months
  • 2

    FDA Approval

    1/8/2024

Other devices from Thirona BV