Intended Use

Intended for head, body, cardiac and vascular X-ray Computed Tomography applications producing cross-sectional images useful for diagnosis and therapy planning.

Technology

Multislice dual-tube CT scanner with 192 detector rows, using a fast scintillator array, direct drive DC motor gantry rotation, and integrated Deep Learning Image Reconstruction (DLIR) software as an alternative image noise reduction method replacing ASIR-CV.

Performance

The device was tested according to relevant IEC standards, verified with phantoms and clinical data demonstrating DLIR achieves equivalent or improved noise reduction and image quality compared to ASIR-CV, without compromising spatial resolution or accuracy. Clinical blinded reader studies confirmed diagnostic image quality is not inferior to predicate devices.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    2/10/2023

    8 months
  • 2

    FDA Approval

    10/13/2023

Ready to Sharpen Your Edge?

Join hundreds of your peers who rely on RadAI Slice. Get the essential weekly briefing that empowers you to navigate the future of radiology.