Intended Use

HipCheck assists the surgeon to determine quantitative measurements for femoroacetabular impingement (FAI) procedures by providing static localization information from intra-operative fluoroscopic images through virtual measurement tools.

Technology

The device processes intra-operative X-ray images, overlays virtual measurement tools on a tablet interface, and includes software compatibility updates such as running on Windows 10. It features an AI/ML model for detecting the femoral head and neck with high accuracy using image processing techniques including Gabor filtering, edge detection, and Hough transformation.

Performance

Performance testing included software verification, bench testing, simulated-use design validation by surgeons, and standalone evaluation of the AI/ML model on 745 fluoroscopic images demonstrating high accuracy in detecting and localizing the femoral head and neck. No clinical or animal testing was performed. Device met electrical and mechanical safety standards.

Predicate Devices

No predicate devices specified