Intended Use

The VX1+ assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia.

Technology

The VX1+ is a software algorithm running on a computing platform that receives analog or digital intra-cardiac signals from electrophysiology recording systems or a 3D mapping system. It performs real-time analysis using machine learning and signal processing to detect dispersed electrograms and annotate them manually or automatically on 3D anatomical maps. It supports bidirectional digital communication allowing automatic tagging with the EnSite X 3D mapping system.

Performance

Non-clinical testing included electromagnetic compatibility, electrical safety, thorough software verification and validation including algorithm module testing, and a reader study validating correlation with expert visual analysis. Usability evaluation in simulated and clinical environments confirmed no safety issues. Clinical testing involved a study with 22 patients confirming reliable detection and auto-tagging of dispersed electrograms without added risk or procedure time.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    11/22/2022

    2 months
  • 2

    FDA Approval

    1/21/2023