Intended Use

Saige-Density is a software application intended for use with compatible full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) systems. Saige-Density provides an ACR BI-RADS Atlas 5th Edition breast density category to aid interpreting physicians in the assessment of breast tissue composition. Saige-Density produces adjunctive information. It is not a diagnostic aid.

Technology

Saige-Density is a Software as a Medical Device that processes screening and diagnostic digital mammograms using deep learning techniques. It inputs 2D DICOM images (FFDM and/or 2D synthetics), producing a study-level breast density category. Outputs are DICOM structured report and secondary capture objects containing the breast density category following BI-RADS guidelines.

Performance

Validation was performed via a multi-site retrospective study on 796 mammogram cases (6,170 images) from five U.S. breast imaging centers distinct from training sites. Accuracy in four-class breast density categorization was 81.28%, with robust statistical analysis by an independent biostatistician.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    7/29/2022

    4 months
  • 2

    FDA Approval

    12/16/2022