Intended Use

The Viz RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements; provides annotated images for clinical review, but results are not intended for standalone clinical decision-making.

Technology

Software-only device using a locked AI/ML algorithm to measure ventricular diameters from CT scans; produces DICOM annotated image series with color overlays indicating ventricular diameters and interventricular septum; hosted on backend server analyzing DICOM images forwarded from CT scanners and exports results to PACS for clinical review.

Performance

Clinical performance testing demonstrated high agreement of AI measurements with ground truth obtained from trained radiologists; mean absolute error was under 7.2 mm, meeting performance goals; performance was consistent across clinical sites, genders, ages, slice thicknesses, and scanner models.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    4/14/2022

    4 months
  • 2

    FDA Approval

    8/29/2022