Intended Use

Noninvasive spot measurement of pulse rate and estimated breathing rate (chest wall movements) using software assessing video footage from a fixed-installation solution in single occupancy rooms within hospitals and secured environments with professional healthcare oversight.

Technology

Software-only medical device that runs on standard off-the-shelf computing hardware with off-the-shelf video cameras and infrared illuminators. It uses proprietary algorithms to extract pulse and breathing rate signals from the acquired video data. The device is designed for spot check measurements when the patient is still.

Performance

Performance testing was conducted through software verification and validation demonstrating measurement accuracy of pulse rate ±3 beats per minute over 50-130 bpm range and estimated breathing rate ±2 breaths per minute over 8-31 bpm range, with clinical evidence from prior clearances supporting safety and effectiveness. The device benefits from extensive clinical validation and human factors/usability assessments ensuring reliable operation in intended environments.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    3/28/2022

    1 month
  • 2

    FDA Approval

    4/29/2022