Intended Use

IDx-DR is indicated for use by healthcare providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. It is indicated for use with the Topcon NW400 fundus camera.

Technology

IDx-DR v2.3 consists of AI software components including IDx-DR Analysis (image processing and classification), IDx-DR Client (user interface and image transfer software), and IDx-DR Service (secure server hosting and cybersecurity). It analyzes fundus images from a Topcon NW400 camera, providing automated diabetic retinopathy diagnosis without other patient data.

Performance

Performance testing including retrospective clinical validation of AI algorithm upgrades demonstrates maintained sensitivity (~88%) and specificity (~90%) in detecting more than mild diabetic retinopathy, with diagnosability over 95% after final image acquisition protocols, supporting substantial equivalence with the predicate device.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    9/21/2021

    8 months
  • 2

    FDA Approval

    6/17/2022