Intended Use

StrokeSENS LVO is a radiological computer-aided triage and notification (CADt) software indicated for use in the analysis of CTA head images to assist hospital networks and trained clinicians in workflow triage by flagging and communication of suspected positive findings of Large Vessel Occlusion (LVO) in head CTA images. It sends notifications to clinicians of suspected LVO that requires review.

Technology

StrokeSENS LVO consists of a DICOM-compliant software system with two main components: a Processing Engine that analyzes CTA head images to detect LVO using a machine learning binary classifier, and a compatible Radiological Software Platform that handles image retrieval and notification workflows. The system outputs a binary positive/negative result and notifies clinicians via email or workstation for triage purposes.

Performance

Performance testing included a retrospective study on 400 independent CTA studies with a balanced prevalence of LVO, achieving 89.4% sensitivity and 87.4% specificity for detecting LVO. The device processes images rapidly, with an average notification time of 0.75 minutes, exceeding the goal of under 5 minutes. Software verification followed ISO and FDA guidelines ensuring safety and effectiveness.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    7/20/2021

    2 months
  • 2

    FDA Approval

    10/14/2021