Intended Use

Indicated to acquire and display cross-sectional volumes of the whole body, including head, with capability to image whole organs in a single rotation; specialized studies by qualified physicians.

Technology

Whole body multislice helical CT scanner using rapid kV switching for spectral imaging; employs AiCE deep neural network noise reduction and FIRST iterative reconstruction to improve image quality and reduce dose.

Performance

Bench and clinical testing show diagnostic quality images with significant noise reduction (up to 32%) and dose reduction (69-81%) compared to prior algorithms; spectral imaging demonstrated equivalent image quality and additional artifact reduction; clinical images reviewed by board-certified radiologists confirmed diagnostic suitability.

Predicate Devices

No predicate devices specified