Intended Use

EchoGo Core is intended to be used for quantification and reporting of results of cardiovascular function to support physician diagnosis in adult populations.

Technology

EchoGo Core is a standalone software that automatically analyzes echocardiogram images to measure ejection fraction, global longitudinal strain, and left ventricular volume using machine learning-based auto-contouring algorithms. It generates quantitative reports based on Simpson's biplane method without manual contour editing.

Performance

The device was validated retrospectively on 378 echocardiogram studies from multiple ultrasound systems demonstrating comparable accuracy to the predicate device TomTec Arena TTA2, with low error rates and eliminated inter-operator variability due to automated contouring. Software verification and cybersecurity testing were also conducted.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    5/1/2019

    6 months
  • 2

    FDA Approval

    11/13/2019

Other devices from Ultromics Ltd