Intended Use

Measurement of R2 and iron concentration in the liver from MRI scans.

Technology

Stand-alone software application using convolutional neural networks for image recognition and analysis of multi-slice, spin-echo MRI data; performs automatic R2 measurement and converts to liver iron concentration using a calibration curve; includes an AI-based image quality control module.

Performance

Performance testing included software verification and validation, user usability testing, repeatability and clinical studies. Clinical study on 971 datasets showed sensitivities and specificities above 90% for clinically relevant liver iron concentration thresholds compared to the predicate device FerriScan. Repeatability was comparable or better than liver biopsy variability.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    8/15/2018

    3 months
  • 2

    FDA Approval

    11/30/2018