Intended Use

Accessory for image-guided interventional and diagnostic procedures involving the prostate gland to enhance visualization of 2D transrectal ultrasound images, enabling segmentation, 3D reconstruction, image review, annotation, and instrument tracking in clinical settings.

Technology

Software that performs 3D reconstruction of TRUS video images, fuses these with pre-acquired 3D images such as MRI or CT, supports segmentation, multiplanar views, image annotation, and records instrument locations. It interfaces with a third-party mechanical stepper device with digital position encoders for TRUS probe tracking, using Windows OS.

Performance

Non-clinical performance testing was conducted including verification and validation of input/output functions. Using synthetic prostate phantoms, image registration accuracy showed a mean alignment error of 2.0 mm and needle placement error of 2.9 mm, deemed acceptable for clinical use.

Predicate Devices

No predicate devices specified

Device Timeline

  • 1

    Submission

    1/27/2017

    4 months
  • 2

    FDA Approval

    6/7/2017