Intended Use

A software aid for notification by analyzing brain CT angiogram images in acute stroke settings to identify suspected large vessel occlusions and notify neurovascular specialists for timely review.

Technology

ContaCT is a software-only system consisting of image forwarding software for transmitting DICOM images, image processing and analysis software hosted on a server that performs vessel segmentation and LVO detection, and a non-diagnostic mobile viewing application for previewing compressed images used for notification purposes.

Performance

In a retrospective study of 300 CT angiogram cases, ContaCT demonstrated 87.8% sensitivity and 89.6% specificity for detecting LVO compared to neuro-radiologist established ground truth. The device significantly reduced notification time to specialists by an average of 51 minutes compared to standard care, potentially enabling earlier intervention.

Device Timeline

  • 1

    Submission

    9/29/2017

    4 months
  • 2

    FDA Approval

    2/13/2018

Ready to Sharpen Your Edge?

Join hundreds of your peers who rely on RadAI Slice. Get the essential weekly briefing that empowers you to navigate the future of radiology.

We respect your privacy. Unsubscribe at any time.